MaxVenturi™OPERATING MANUAL & INSTRUCTIONS FOR USER211P03R211M03 Rev.J 201030405060ABCDEF

Strona 2

866.4.Maxtec www.maxtecinc.com42.2 CalibrationA new calibration is required when:• The measured O2 percentage in 100% O2 is below 97.0%.• The mea

Strona 3 - CLASSIFICATION

866.4.Maxtec www.maxtecinc.com 52.2.2 100% Oxygen CalibrationNote: Analyzer will read "Cal Err St" if the sample gas is not stable or if

Strona 4 - WARNINGS

866.4.Maxtec www.maxtecinc.com62.3 Device Set-upNote: Patient circuit and patient interface should be assembled according to manufacturers instruct

Strona 5 - ATTENTION:

866.4.Maxtec www.maxtecinc.com 72.4 Flow and Oxygen Settings Adjustment2.4.1 Initial Settings2.4.2 Changing Flow SettingsAdjust flow to desired sett

Strona 6 - TABLE OF CONTENTS

866.4.Maxtec www.maxtecinc.com82.4.3 Changing Oxygen Settings2.5 DisposablesThe MaxVenturi is intended for use with approved disposables. Several p

Strona 7 - 1.0 SYSTEM OVERVIEW

866.4.Maxtec www.maxtecinc.com 9The delivery circuits can be paired with the patient interface circtuits listed in the table below: Table Instruct

Strona 8

866.4.Maxtec www.maxtecinc.com10 3.0 FACTORS INFLUENCING ACCURATE READINGS3.1 Elevation/Pressure Changes • Changes in elevation result in a readin

Strona 9 - 2.0 SET-UP INSTRUCTIONS

866.4.Maxtec www.maxtecinc.com 11CAL Err hi: SENSOR VOLTAGE TOO HIGH• Repeat calibration routine as described in this manual. If unit repeats this

Strona 10 - 2.2.1 Room Air Calibration

866.4.Maxtec www.maxtecinc.com12 6.0 CHANGING THE OXYGEN SENSORTurn off analyzer. Disconnect sensor from sensor cable.Remove sensor from sensor por

Strona 11 - 2.2.2 100% Oxygen Calibration

866.4.Maxtec www.maxtecinc.com 13 8.0 SPECIFICATIONS8.1 Analyzer SpecificationsMeasurement Range:...

Strona 13 - 2.4.2 Changing Flow Settings

866.4.Maxtec www.maxtecinc.com148.3 Oxygen Inlet Supply:This device was designed for an oxygen inlet pressure of 3.5 bar (50psi). Certain geographi

Strona 14 - 2.5 Disposables

866.4.Maxtec www.maxtecinc.com 15 10.0 TROUBLESHOOTINGProblem: The flow does not reach 55LPM even with the flow valve wide open.Possible Cause: The p

Strona 15 - 2.5.3 Inlet Filter:

866.4.Maxtec www.maxtecinc.com I CLASSIFICATIONClassification:...Class II medical deviceProtection again

Strona 16 - 3.2 Temperature Effects

866.4.Maxtec www.maxtecinc.comII WARNINGS Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.• T

Strona 17 - 5.0 CHANGING THE BATTERIES

866.4.Maxtec www.maxtecinc.com III CAUTION: Indicates a potentially hazardous situation, if not avoided, could result in minor or moderate injury a

Strona 18 - 7.0 CLEANING AND MAINTENANCE

866.4.Maxtec www.maxtecinc.comIV TABLE OF CONTENTSClassification ... IWarranty ...

Strona 19 - 8.0 SPECIFICATIONS

866.4.Maxtec www.maxtecinc.com 1 1.0 SYSTEM OVERVIEW1.1 Component IdentificationON/OFF Button:Calibration Button:MAX-250 Series Oxygen SensorLow Bat

Strona 20

866.4.Maxtec www.maxtecinc.com21.2 Symbol GuideThe following symbols and safety labels are found on the MaxVenturi:!LPM%Attention, consult accompan

Strona 21 - 10.0 TROUBLESHOOTING

866.4.Maxtec www.maxtecinc.com 32...30-90 minutes346Remove Max-250 sensor from package and pull off barrier film. Wait 30 to 90 minutes for sensor t