MaxVenturi™OPERATING MANUAL & INSTRUCTIONS FOR USER211P03R211M03 Rev.J 201030405060ABCDEF
866.4.Maxtec www.maxtecinc.com42.2 CalibrationA new calibration is required when:• The measured O2 percentage in 100% O2 is below 97.0%.• The mea
866.4.Maxtec www.maxtecinc.com 52.2.2 100% Oxygen CalibrationNote: Analyzer will read "Cal Err St" if the sample gas is not stable or if
866.4.Maxtec www.maxtecinc.com62.3 Device Set-upNote: Patient circuit and patient interface should be assembled according to manufacturers instruct
866.4.Maxtec www.maxtecinc.com 72.4 Flow and Oxygen Settings Adjustment2.4.1 Initial Settings2.4.2 Changing Flow SettingsAdjust flow to desired sett
866.4.Maxtec www.maxtecinc.com82.4.3 Changing Oxygen Settings2.5 DisposablesThe MaxVenturi is intended for use with approved disposables. Several p
866.4.Maxtec www.maxtecinc.com 9The delivery circuits can be paired with the patient interface circtuits listed in the table below: Table Instruct
866.4.Maxtec www.maxtecinc.com10 3.0 FACTORS INFLUENCING ACCURATE READINGS3.1 Elevation/Pressure Changes • Changes in elevation result in a readin
866.4.Maxtec www.maxtecinc.com 11CAL Err hi: SENSOR VOLTAGE TOO HIGH• Repeat calibration routine as described in this manual. If unit repeats this
866.4.Maxtec www.maxtecinc.com12 6.0 CHANGING THE OXYGEN SENSORTurn off analyzer. Disconnect sensor from sensor cable.Remove sensor from sensor por
866.4.Maxtec www.maxtecinc.com 13 8.0 SPECIFICATIONS8.1 Analyzer SpecificationsMeasurement Range:...
866.4.Maxtec www.maxtecinc.com148.3 Oxygen Inlet Supply:This device was designed for an oxygen inlet pressure of 3.5 bar (50psi). Certain geographi
866.4.Maxtec www.maxtecinc.com 15 10.0 TROUBLESHOOTINGProblem: The flow does not reach 55LPM even with the flow valve wide open.Possible Cause: The p
866.4.Maxtec www.maxtecinc.com I CLASSIFICATIONClassification:...Class II medical deviceProtection again
866.4.Maxtec www.maxtecinc.comII WARNINGS Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.• T
866.4.Maxtec www.maxtecinc.com III CAUTION: Indicates a potentially hazardous situation, if not avoided, could result in minor or moderate injury a
866.4.Maxtec www.maxtecinc.comIV TABLE OF CONTENTSClassification ... IWarranty ...
866.4.Maxtec www.maxtecinc.com 1 1.0 SYSTEM OVERVIEW1.1 Component IdentificationON/OFF Button:Calibration Button:MAX-250 Series Oxygen SensorLow Bat
866.4.Maxtec www.maxtecinc.com21.2 Symbol GuideThe following symbols and safety labels are found on the MaxVenturi:!LPM%Attention, consult accompan
866.4.Maxtec www.maxtecinc.com 32...30-90 minutes346Remove Max-250 sensor from package and pull off barrier film. Wait 30 to 90 minutes for sensor t